Handmade Air Geiger Counter

Operation voltage feature

The power supply voltage Vpp is changed and operation voltage is measured.
The high voltage of 3000V to 5000V cannot be measured by high impedance, the special measurement circuit is needed.

Vpp voltage vs. operation voltage to GM tube
Vpp voltage[V]A/DOperation voltage X[V]
51.38/2822256
61.67/3432744
71.98/4053240
82.29/4683744
92.58/5294232
102.84/5814648
113.09/6335064


It is changing linearly.
The analog voltage A and the operation voltage X can be drawn from the following expression of relations. At last, operatin voltage X is
X[V]alpha x6.5 xD
If digital value D is known, the operation voltage X is known.
In order to simplify the digital calculation by PIC, it is referred to as X[V] =8xD. If it is 8 times, only shift operation is required.
75K has counted backward as divided resistance.(To be set to alpha x 6.5 = 8).
Although there are some errors (several 10 V), it is used enough.
In addition, the resolution of the above-mentioned thing to operation voltage is 8V.